ET-100 Handheld Thermal Emissometer
- In-Band Reflectance & Emittance
- Easy to Use
- ASTM Compliant
- Dual Angles of Incidence
- Save Time
- Room Temperature Samples
Total Hemispherical Emittance Measurements
The ET-100 measures directional reflectance from 1.5 to 21 μm and, based on those values, calculates directional and total hemispherical emissivity. The ET-100 emissometer conforms to ASTM E408, the standard test method for the determination of emittance using a portable instrument. In-band reflectance data for six discrete bands.
|
Measured Parameter |
Directional hemispherical reflectance (DHR) |
|
Method |
Integrated total reflectance in a band for a given angle of incidence |
|
Measured Value |
Absolute reflectance (0-1) |
|
Calculated Value |
Directional thermal emissivity at 20º, directional thermal emissivity at 60º, |
|
Wavelength Bands (microns) |
1.5-2.0, 2.0-3.5, 3.0-4.0, 4.0-5.0, 5.0-10.5, 10.5-21 |
|
Angle of Incidence |
20º & 60º from normal incidence |
|
Calibration Coupon |
Specular Gold |
|
Accuracy |
+/- .03 |
|
Repeatability |
± .005 units |
|
Beam Spot Size |
0.50 inches |
|
Measurement Time |
10 sec |
|
Sample Size & Geometry |
Flat: ≥ 0.5 in. diameter Curved: 6 in. convex; 12 in. concave |
|
Warm Up Time |
90 seconds |
|
Time Between Measurements |
2 seconds |
|
Sample Temperature |
Ambient or heated/cooled to 0 – 100º C |
|
Operating Temp |
0º to 40º C |
|
Storage Temp |
-25º to 70ºC |
|
Run Time |
2 hours on one battery. Battery easily replaced with continuous operation after battery replacement. |
|
Power Source |
Rechargeable battery (standard environmentally friendly NiMH) |
|
Battery Recharge Time |
1 hour |
|
IR Source |
Kanthal filament operated at about 1,000ºC |
|
Weight |
4.7 lb |
|
Form Factor |
H 11.54”, L 9.04”, W 3.27” (29.31 cm x 22.96 cm x 9.44 cm) |
|
Operator Interface |
LCD graphics screen, 1/4 VGA, touch screen, software buttons; trigger switch in handle |
|
Software Applications |
Pass/fail can be incorporated, user set values |
|
Diagnostics |
On screen status and signals monitor. Signal values stored with data. Raw data collection and display. |
|
Data Format |
Data files can be opened and post processed with Excel or a text processor |
|
Data Storage |
Removable SanDisk (SD) card |
|
ASTM Standard |
E408 |
|
Export Control |
ECCN #3A999.F |